Keypad Display Information
Press the f1, f2 and f3 function keys to perform the action shown above each key on the display.
Press to turn the meter on.
When the meter is on, press and quickly release to turn the display backlight on or off or hold down to turn the meter off.
In the measurement mode, press to take a measurement.
In the setup, calibration and other menus, press to escape the current menu and return to the measurement mode.
In the measurement mode, press to enter the setup menu.
In the setup, calibration and other menus, press to scroll up through a list of options.In the continuous measurement mode, press to freeze the displayed measurement and press again to unfreeze the measurement.
In the setup, calibration and other menus, press to scroll left through a list of options. In the measurement mode, press to change the displayed measurement mode. Options are pH, mV, RmV (relative mV) and ORP.
In the setup, calibration and other menus, press to scroll right through a list of options.In the measurement mode, press to log or print a measurement. In the setup, calibration and other menus, press to scroll down through a list of options.Press to view the data log and calibration log.
Press to start or stop the stirrer probe.
Keypad
1. Press to power the meter on. When the meter is on, press and quickly release to turn the
backlight on or off or press and hold for about three seconds to power the meter off.2. Press to exit any meter function and return to the measurement mode. 3. The f1, f2, and f3 function keys perform a variety of meter operations. The menu-specific operation is
shown above each key. For example, press f1 in the measurement mode to start a calibration.4. The , , or keys are used as navigation keys (up, right, down, left) when selecting from a fixed list or grid of meter options. In the measurement mode, these keys are used to access the setup menu, change the measurement mode, manually log or print a measurement and hold (freeze) a displayed measurement. 5. Press to turn on or off the stirrer probe (Cat. No. 096019). 6. Press to access the calibration log and data log.3
pH Calibration
One to five pH buffers can be used for calibration. Always use fresh pH buffers and select buffers that bracket the sample pH and are one to four pH units apart. Prepare the pH electrode according to the instructions in the electrode use guide. Connect the pH electrode and any other electrodes to be used (ATC probe, stirrer probe, reference electrode) to the meter. Power on the meter and set the measurement mode to pH.1. In the measurement mode, press f1 (cal).
2. Rinse the pH electrode and any other electrodes in use with distilled water, blot dry with a lint-free tissue and place into the pH buffer.3. When the electrode and buffer are ready, press f3 (start).
4. Wait for the pH value on the meter to stabilize and stop flashing and perform one of the following actions:a. Press f2 (accept) to accept the displayed pH value.
b. Press f3 (edit) to access the numeric entry screen and edit the pH buffer value.i. Press , , or to highlight a number, decimal point or negative sign; press f3 (enter) to select the highlighted item and repeat until the buffer value at the measured temperature is shown above the numeric entry screen. ii. Press f2 (done) to exit the numeric entry screen. iii. Press f2 (accept) to accept the entered pH value.
5. Press f2 (next) to proceed to the next buffer and repeat steps 2 through 4 or press f3 (cal done) to save and end the calibration. If five buffers are used, the calibration will save and end once the fifth buffer value is accepted.a. If a one point calibration is performed, press f2 (accept) to accept the displayed slope value or press f3 (edit) to access the numeric entry screen, enter the slope value and press f2 (accept). 6. The meter will display the calibration summary including the average slope. Press f1 (meas) to export the data to the calibration log or press f2 (print) to export the data to the calibration log and a printer or computer. The meter will automatically proceed to the measurement mode.
4
Measurement
Press while taking a measurement in the continuous measurement mode to freeze the display and press a second time to unfreeze the display and continue the measurement. Press while taking a measurement to manually export the measurement to the data log, if the data log is enabled in the setup menu.
1. Rinse the pH electrode and any other electrodes in use with distilled water, blot dry with a lint-free tissue and place into the sample.2. If the stirrer probe is in use, press to turn on the stirrer probe. 3. Start the measurement and wait for it to stabilize.
a. If the meter is in AUTO-READ mode (default setting), press to start the measurement. When the
icon stops flashing, record the pH and temperature of the sample. Press again to start a new measurement.b. If the meter is in continuous mode, the meter will immediately start taking a measurement and update the display whenever the measurement changes. Wait for the display to show ready and record the pH and temperature of the sample.c. If the meter is in timed mode, the meter will log measurements at the preselected time interval, regardless of the measurement stability. The meter will update the display whenever the
measurement changes, so the pH and temperature of the sample can be recorded when the display shows ready.4. If the stirrer probe is in use, press to turn off the stirrer probe.5. Remove the electrode from the sample, rinse with distilled water, blot dry and place into the next sample.6. Repeat steps 2 through 5 for all samples.
7. When all samples have been measured, store the electrode according to its user guide.
5
Setup Menu
Navigating the Setup Menu1. In the measurement mode, press to enter the main setup menu. , or to scroll through the main setup menu options. Press f3 (select) to 2. Press , select a main setup menu option. 3. Press or to scroll through setup submenu options. Press f3 (select) to select a setup submenu option.4. Perform the appropriate actions to set the desired parameter in the setup submenus.
a. To select a value from a list of options, press f3 (select) to set the value.
or to highlight the desired value and press b. To enter a numeric value, use the numeric entry screen.
i. Select the value to be entered by pressing f3 (select) or f3 (edit). The numeric entry screen will popup on the display.ii. Press , , or to highlight a number, decimal place or negative sign; press f3 (enter) to select the highlighted item and repeat until the desired value is shown on the top of the numeric entry screen.iii. Press f2 (done) to save the value and exit the numeric entry screen.5. Press f1 (back) and then to return to the measurement mode at any time.Setup Menu Overview
pH ChannelMethodMode & Settingst Measure ModetRead TypetResolutiontBuffer GrouptStabilitytAveragingtAlarm SettingstSample ID Settingst Export Datat Data Logt Date / Timet Languaget Key Press Beept Alarm Beept Stirrert Contrastt Auto Shut Offt User ID.Log Viewt Data Logt Calibration LogDiagnosticst Meter Self Testt Factory Resett About MeterTemperaturetManual Temp ValuetTemperature UnittTemperature Calibration6
pH Buffer Group Selection
The selected buffer group allows for the automatic recognition of certain pH buffers during a pH calibration. The USA buffer group includes pH 1.68, 4.01, 7.00, 10.01 and 12.46 buffers and the DIN buffer group includes pH 1.68, 4.01, 6.86, and 9.18 buffers.1. In the measurement mode, press 2. Press , , or to highlight pH Channel and press f3 (select).
3. Press or to highlight Mode and Settings and press f3 (select).4. Press or to highlight Buffer Group and press f3 (select).5. Press or keys to highlight USA or DIN and press f3 (select).
6. Press to return to the measurement mode.Read Type Selection
1. In the measurement mode, press . 2. Press , , or to highlight pH Channel and press f3 (select).
3. Press or to highlight Mode and Settings and press f3 (select).4. Press or to highlight Read Type and press f3 (select).
5. Press or to highlight Auto, Continuous or Timed and press f3 (select).
a. If Timed is selected and the time interval needs to be changed – highlight Timed; press to highlight hours (HH), minutes (MM) or seconds (SS); press f3 (edit) to access the numeric entry screen; use the numeric entry screen to change the values and press f1 (back) when the time interval is correct. 6. Press to return to the measurement mode.7
Viewing the Calibration Log
1. In the measurement mode, press 2. Press 3. Press or or . to highlight Calibration Log and press f2 (accept). to highlight pH – Channel and press f2 (select).
to highlight pH, RmV, ORP or Temperature as the calibration type and 4. Press or press f2 (select).5. The meter will display a list of calibrations for the selected channel and calibration type. The list shows the sequential number of the calibration and the date and time it was saved (07/01/2011 12:45). or to highlight a specific calibration and press 6. To view the calibration data, press f2 (select). Press f2 (print) to print the calibration, press f3 (info) to view the electrode slope between calibration points or press f1 (back) to return to the list of calibrations.7. Press to return to the measurement mode. Viewing the Data Log
1. In the measurement mode, press 2. Press or . to highlight Data Log and press (accept).
3. The meter will display a list of the data points. The list shows the sequential number of the data point and the date and time the data point was saved (07/01/2011 12:45). or to highlight the data 4. To view the measurement information for an individual data point, press point and press f2 (select). Press f2 (print) to print the data point or press f1 (back) to return to the list of data points.5. Press to return to the measurement mode. Water Analysis Instruments www.thermoscientific.com/waterNorth America 166 Cummings Center Beverly, MA 01915 USA Toll Free: 1-800-225-1480 Tel: 1-978-232-6000 info.water@thermo.com©2011 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. & its subsidiaries.